A formed flange that flexes under bolt load loses sealing contact; set flange width, return geometry and boss support before the housing is released.
What the failure looks like on the line
Equipment housings for clinical environments are judged on cleanability, fit and surface finish, with documentation reviewed separately from the part itself.
The perimeter flange of the housing deflects when the cover bolts are tightened, and the sealing line lifts between fasteners.
A thin, flat flange behaves as a cantilever plate and bends under local bolt load, so compression is concentrated at the fixing points and lost between them. A return lip or a bonded boss converts that bending into much stiffer edge stiffness.
Start with the failure as it is reported, not with the parameter that is easiest to measure. On a buyer's specification this distinction matters: a part can measure inside tolerance and still fail in use, because the drawing records geometry while the complaint records behaviour. Flange width is usually the first value both sides compare, and it is also the one most often quoted without the conditions attached to it.
The process window that decides it
Four variables move this result more than the rest. The ranges below are illustrative engineering guidance for a first discussion; confirm them on your own tool, material grade and part geometry before they become contract terms.
| Variável | Typical window | What fails outside it |
|---|---|---|
| Flange width | At least 15-25 mm | A narrow flange bends noticeably under bolt load |
| Flange sheet gauge after forming | Not less than 70 % of nominal | Thinned flange loses bending stiffness quickly |
| Return lip depth | 8-15 mm where the design allows | Without a return the flange stays a flat cantilever |
| Fastener spacing | 80-120 mm along the sealing line | Wider spacing leaves unsupported spans that lift |
Table 1. Starting window for a joint process trial. Values are engineering guidance, not a measured survey.
Flange width is typically held at At least 15-25 mm. A narrow flange bends noticeably under bolt load Record it together with the measurement point and the moment in the cycle; a single number without a location is not a process setting.
Flange sheet gauge after forming is typically held at Not less than 70 % of nominal. Thinned flange loses bending stiffness quickly When this drifts, the symptom usually appears at one cavity, one sheet position or one lane before it appears across the whole batch.
Return lip depth is typically held at 8-15 mm where the design allows. Without a return the flange stays a flat cantilever A change here is cheap during a trial and expensive after tooling release, so treat it as a controlled variable with a documented re-approval.
Fastener spacing is typically held at 80-120 mm along the sealing line. Wider spacing leaves unsupported spans that lift Put it on the change-control list, otherwise an unannounced material or setting substitution will be mistaken for ordinary process noise.
Reading the response curve

The left panel shows rim deflection under load (0.1 mm) as flange width (mm) increases: the response falls from 82 to 12 on the illustrative 0–100 scale. The dashed line at 30 marks a proposed acceptance limit, not a universal standard. Move that line only after the test method and the sample plan have been agreed, otherwise two suppliers can both be compliant and still deliver different parts.
Read the shape rather than a single point. A curve that changes quickly near the limit is more sensitive to ordinary process noise than one that approaches it gradually, and that difference should be reflected in how tightly the setting is controlled and how often it is checked.
Where the causes actually sit
The right panel of Figure 1 splits audited failures of this type into five groups. The percentages are illustrative engineering guidance for prioritising an investigation, not a measured survey and not a prediction for your line.
| Causa | Illustrative share | First check |
|---|---|---|
| Flange too narrow | 32% | Measure the formed flange width and wall thickness at several points around the perimeter. |
| Flange thinned in forming | 25% | Check the flange for a return lip or bonded boss at each fixing point. |
| No return lip or boss | 19% | Apply the specified bolt torque and measure rim deflection between fasteners. |
| Fastener spacing too wide | 14% | Inspect the sealing line for continuous contact using an agreed indicator method. |
| Bolt torque too high | 10% | Record fastener spacing against the approved drawing before sign-off. |
Table 2. Cause split and the corresponding first check. Shares are indicative and sum to 100.
Flange too narrow accounts for roughly 32% of the illustrative split. Measure the formed flange width and wall thickness at several points around the perimeter. Start here before authorising a material or tooling change, because it is the fastest explanation to confirm or eliminate.
Flange thinned in forming accounts for roughly 25% of the illustrative split. Check the flange for a return lip or bonded boss at each fixing point. Start here before authorising a material or tooling change, because it is the fastest explanation to confirm or eliminate.
No return lip or boss accounts for roughly 19% of the illustrative split. Apply the specified bolt torque and measure rim deflection between fasteners. Start here before authorising a material or tooling change, because it is the fastest explanation to confirm or eliminate.
Checks before sign-off
In practice the sequence below is cheaper to run than a root-cause debate after a rejected delivery.
- Measure the formed flange width and wall thickness at several points around the perimeter.
- Check the flange for a return lip or bonded boss at each fixing point.
- Apply the specified bolt torque and measure rim deflection between fasteners.
- Inspect the sealing line for continuous contact using an agreed indicator method.
- Record fastener spacing against the approved drawing before sign-off.
Keep a retained reference sample beside every result. When a batch is disputed, the comparison that settles it is almost always between the retained part and the current one under the same method, not between either part and a number on a drawing. Record the instrument, the operator, the conditioning time and the position on the part or sheet for each reading.
Writing it into the specification and the evidence pack
Rim deflection between fasteners measured at the specified bolt torque shall remain within the agreed limit, with flange width and wall thickness recorded at the stated measuring points.
Write the requirement as a behaviour at a stated condition, not as a brand of material or a machine setting. State the test method, the conditioning, the sample plan and the rule for deciding a pass. Where a document is needed, ask the supplier for documentation that covers the actual construction and check its scope against your own market before approval; a document for the base material alone does not describe the finished part you are buying.
At first-article approval, freeze the drawing revision, the material and colour identification, the process window, the tool and cavity identification, and the inspection method. Any later change to grade, site, tool, or window should trigger a new first article rather than a paper waiver. Keep retained samples from the approved lot and from each production lot so that a future comparison is possible.
| Evidence item | What it must show |
|---|---|
| Approved drawing and revision | The exact geometry and tolerances the decision applies to |
| Material and colour identification | Grade, supplier, colour reference and any declared recycled content |
| Process window and change log | The agreed settings and any deviation since approval |
| Inspection record and retained samples | Method, conditions, results and a physical reference from the same lot |
Table 3. Minimum evidence pack for a first-article decision and a later batch dispute.
None of the figures above should be read as a measured survey. They are engineering guidance for framing a joint trial; confirm them on your own tool, grade and geometry before they become contract terms.
Questions buyers ask
Is this a material problem or a process problem?
A thin, flat flange behaves as a cantilever plate and bends under local bolt load, so compression is concentrated at the fixing points and lost between them. A return lip or a bonded boss converts that bending into much stiffer edge stiffness. It is worth separating the two before a supplier changes a grade: hold the material constant for one controlled run, then move one setting at a time. If the symptom follows the setting, the material was never the cause.
Can flange width be traded against cycle time?
Only within the window stated in Table 1, and only with a recorded trial. A narrow flange bends noticeably under bolt load A faster cycle that sits outside the window usually moves the failure downstream, where it costs more to find.
What should be recorded at incoming inspection?
Keep the measurement, the method, the sample identity and the conditions together. Measure the formed flange width and wall thickness at several points around the perimeter. A result without its conditions cannot be compared with a later batch.
Does a change of medical device require re-approval?
Treat any change to material grade, tool, site or process window as a change that needs a new first-article decision. For thick-gauge forming work the drawing alone is not enough; the process route is part of what was approved.
How many samples are enough for a first-article decision?
There is no universal number. Sample across cavities, lanes, sheet positions or reel lengths so that the spread is visible, not just the average. Agree the sample plan and the acceptance rule with the supplier before the trial starts.
What belongs in the evidence pack?
The approved drawing and revision, the material and colour identification, the process window, the inspection method and results, the retained reference samples, and documentation covering the actual construction. Ask the supplier to state the scope of each document and check it against your own market.
Seeing this on a current thick-gauge forming job?
Send the drawing, the material and grade, the process window you are running, and the point in the cycle or the shelf life where the fault appears. We will help you separate a material question from a process question and a specification question before anyone changes a tool.
Continue reading
- Rib Design in Thermoformed Parts: Stiffness Without a Thin Corner
- Thermoforming Undercuts: Forming a Feature That Has to Release
- Joining and Assembling Thermoformed Parts: Adhesives, Welds and Fasteners
- Leak Testing Thermoformed Containers: Matching the Test to the Leak
- A gasket that will not seat: groove dimensions on a window profile at goods-in